Addressing the key industry challenge of testing 2D materials, which are highly sensitive to water, oxygen, and contaminants, this system has been optimized through extensive R&D and validation. Equipped with specialized accessories for 2D materials, it delivers high-precision temperature control across the full operating range:Within a temperature range of -190°C to 300°C, samples exhibit no frost formation or discoloration when observed under a 50× optical microscope, effectively resolving a critical bottleneck in 2D material testing.
Cool-down Curve: RT ~ -190°C under Vacuum Environment
Heat-up Curve: RT ~ 400°C under Vacuum Environment
1.Cooling & Heating Method: Liquid nitrogen cooling, resistive heating
2.Temperature Control Range: -190°C ~ 600°C (-190°C ~ 400°C under vacuum)
3.Temperature Stability: ±0.1°C (±0.3°C at -120°C ~ -190°C)
4.Temperature Ramp Rate:Max heating rate: 150°C/min;Max cooling rate: 40°C/min
5.Optical Path: Reflective
6.Viewport Size: φ25mm*1mm
7.Viewport Material: JGS2 fused silica (transmission range: 220nm ~ 2500nm), manually removable and replaceable
8.Viewport Defrosting: Equipped with a gas purge bracket; defrosting via liquid nitrogen exhaust gas purge at low temperatures
9.Distance from Viewport Upper Surface to Sample Stage: 7.5mm
10.Sample Chamber Height: 6.5mm
11.Probes: 4*fixed probe holders + 4*tungsten carbide gold-plated probes
12.Probe Interfaces: 4*BNC
13.Sample Stage Potential: Grounded
14.Sample Stage: Silver stage, 23mm*23mm
15.Chamber: Vacuum chamber; mechanical pump vacuum level ≤ 1 Pa